Ad Code

Responsive Advertisement

Ticker

6/recent/ticker-posts

UV SPECTROPHOTOMETER pdf

          UV SPECTROPHOTOMETER
Introduction 
 Ultraviolet-visible (UV-Vis) spectroscopy is one of the most popular 
analytical techniques because it is very versatile and able to detect nearly every molecule. 
With UV-Vis spectroscopy, the UV-Vis light is passed through a sample and the 
transmittance of light by a sample is measured. From the transmittance (T), the absorbance 
can be calculated as A=-log (T). An absorbance spectrum is obtained that shows the 
absorbance of a compound at different wavelengths. The amount of absorbance at any 
wavelength is due to the chemical structure of the molecule. 
 UV-Vis can be used in a qualitative manner, to identify functional groups or 
confirm the identity of a compound by matching the absorbance spectrum. It can also be 
used in a quantitative manner, as concentration of the analyte is related to the absorbance 
using Beer'sLaw. UV-Vis spectroscopy is used to quantify the amount of DNA or protein in 
a sample, for water analysis, and as a detector for many types of chromatography. Kinetics 
of chemical reactions is also measured with UV-Vis spectroscopy by taking repeated UVVis measurements over time. UVVis measurements are generally taken with a 
spectrophotometer. UV-Vis is also a very popular detector for other analytical techniques, 
such as chromatography, because it can detect many compounds. 
Principle 
 UV-Vis is often called a general technique because most molecules will absorb in 
the UV-Vis wavelength range. The UV extends from 100–400 nm and the visible spectrum 
from 400–700 nm. The 100–200 nm range is called the deep UV. Light sources are more 
difficult to find for this range, so it is not routinely used for UV-Vis measurements. Typical 
UV-Vis spectrometers use a deuterium lamp for the UV that produces light from 170–375 
nm and a tungsten filament lamp for visible, which produces light from 350–2,500 nm. 
When a photon hits a molecule and is absorbed, the molecule is promoted into a 
more excited energetic state. UV-visible light has enough energy to promote electrons to a 
higher electronic state, from the highest occupied molecular orbital (HOMO) to the lowest 
unoccupied molecular orbital (LUMO). The energy difference between the HOMO and the 
LUMO is called the band gap. Typically, these orbitals are called bonding and anti-bonding. 
The energy of the photon must exactly match the band gap for the photon to be absorbed. 
Thus, molecules with different chemical structures have different energy band gaps and 
different absorption spectra. 
The most common transitions that fall in the UV-Vis range are π-π* and n- π*. Pi orbitals 
arise due to double bonds, and n- orbitals are for non-bonding electrons. Pi star are antibonding pi orbitals. Thus, the best UV-Vis absorption is by molecules that contain double 
bonds. Pi orbitals adjacent to each other that are connected, called conjugation, typically 
increases absorption. 
Sigma-σ* transitions, associated with single bonds, are higher energy and fall in the deep 
UV, so they are less useful for routine use. The appearance of broad bands or shoulders on 
the UV-Vis structure is due to the numerous vibrational and rotational states of a molecule, 
which lead to separate energy band gaps of slightly different energies. 
 For molecules with absorption in the visible region, the compounds will often appear 
colored. 
However, a common misconception is that the wavelength of peak absorption (λmax) for a 
compound is the color it appears. A compound that appears red does not have much 
absorption in the red region of the spectrum. Instead, the λmax for a compound that looks red 
is green. The color of a compound arises because those wavelengths of light are selectively 
transmitted through the sample, and thus they are not absorbed. A color wheel is helpful in 
determining what color a compound will absorb and what range the λmax will be, as the color 
directly across the wheel from the observed color is the color that is most absorbed. 
Absorption follows Beer's Law, A= εbC where ε is the molar attenuation coefficient, b is 
path length, and C is concentration. The molar attenuation coefficient is the characteristic of 
an individual compound to absorb at a given wavelength and this property is due to 
functional groups, conjugation, etc. If a compound does not have a high attenuation coefficient, it could be tagged with an appropriate group to increase its absorbance. Path 
length is generally related to the size of the cuvette and is 1 cm in standard 
spectrophotometers. 
 UV-Vis is performed on a variety of instruments, from traditional spectrophotometers 
to more modern-day plate readers. The absorbance wavelength must be chosen, either using 
a filter or a monochromator. A monochromator is a device that separates the wavelengths of 
light spatially and then places an exit slit where the desired wavelength of light is. 
Monochromators can be scanned to provide a whole absorbance spectrum. Alternatively, a 
diode-array instrument allows all colors of light to be transmitted through the sample, and 
then the light is separated into different wavelengths spatially and detected using 
photodiodes. Diode-array instruments collect full spectra faster, but are more complicated 
and more expensive. 
Working Procedure 
1. Calibrate the Spectrometer 
• Turn on the UV-Vis spectrometer and allow the lamps to warm up for an 
appropriate period of time (around 20 min) to stabilize them. 
• Fill a cuvette with the solvent for the sample and make sure the outside is clean. 
This will serve as a blank and help account for light losses due to scattering or 
absorption by the solvent. 
• Place the cuvette in the spectrometer. Make sure to align the cuvette properly, as 
often the cuvette has two sides, which are meant for handling (may be grooved) 
and are not meant to shine light through. 
• Take a reading for the blank. The absorbance should be minimal, but any 
absorbance should be subtracted out from future samples. Some instruments might 
store the blank data and perform the subtraction automatically. 
2. Perform an Absorbance Spectrum 
• Fill the cuvette with the sample. To make sure the transfer is quantitative, rinse 
the cuvette twice with the sample and then fill it about ¾ full. Make sure the 
outside is clean of any fingerprints, etc. 
• Place the cuvette in the spectrometer in the correct direction. 
• Cover the cuvette to prevent any ambient light. 
• Collect an absorbance spectrum by allowing the instrument to scan through 
different wavelengths and collect the absorbance. The wavelength range can be 
set with information about the specific sample, but a range of 200–800 nm is 
standard. A diodearray instrument is able to collect an entire absorbance spectrum 
in one run. 
• From the collected absorbance spectrum, determine the absorbance maximum 
(λmax). Repeat the collection of spectra to get an estimate of error in λmax. 
• To make a calibration curve, collect the UV-Vis spectrum of a variety of different 
concentration samples. Spectrometers are often limited in linear range and will not 
be able to measure an absorbance value greater than 1.5. If the absorbance values 
for the sample are outside the instrument's linear range, dilute the sample to get 
the values within the linear range. 
3. Kinetics Experiments with UV-Vis Spectroscopy 
• UV-Vis can be used for kinetics experiments by examining the change in 
absorbance over time. For a kinetics experiment, take an initial reading of the 
sample 
• Quickly add the reagent to start the chemical reaction 
• Stir it well to mix with the sample. If a small amount is added, this could be done 
in a cuvette. Alternatively, mix the reagent with sample and quickly pour some in 
a cuvette for a measurement 
• Measure the absorbance at the λmax for the analyte of interest over time. If using 
up the reagent being measuring (i.e. absorbance is going up because there is less 
reagent to absorb), then the decay will indicate the order of the reaction 
• Using a calibration curve, make a plot of analyte concentration vs time, converting 
the absorbance value into concentration. From there, this graph can be fit with 
appropriate equations to determine the reaction rate constants
Applications 
1. UV-Vis is used in many chemical analyses. It is used to quantitate the amount of 
protein in a solution, as most proteins absorb strongly at 280 nm 
2. UV-Vis is also used as a standard technique to quantify the amount of DNA in a 
sample, as all the bases absorb strongly at 260 nm 
3. RNA and proteins also absorb at 260 nm, so absorbance at other wavelengths can be 
measured to check for interferences. Specifically, proteins absorb strongly at 280 nm, so 
the ratio of absorbance at 280/260 can give a measure of the ratio of protein to DNA in a 
sample 

Post a Comment

0 Comments